Troubleshooting Table

If you already have the hardware and something is not working, start with this table. Most problems fall into wiring, electrodes, power, acquisition, or Wireless pairing.

Symptom Most likely cause First checks
Flat line No EMG output is being read, ADC wiring is wrong, or electrodes do not form a path Power, S/G, ADC input, electrode contact.
Saturated waveform Floating electrodes, bad input contact, wrong ADC range Improve electrode contact and confirm ADC range covers 0–3.3 V.
Large noise Bad electrode contact, laptop power noise, cable motion Stabilize electrode contact, use battery power, and move away from displays/adapters.
No difference between relaxed and clenched Wrong muscle location, unstable pressure, weak activation Re-place IN+ / IN- along the target forearm muscle.
Flex does not show up over USB Flex is not a USB acquisition device Use ADC, MCU, or oscilloscope.
ESP32 values are unstable ESP32 ADC and power-noise limits Use for validation; use external ADC or STM32 ADC + DMA for stability.
Wireless has no data Dongle not connected, Edge off, not paired, or low battery Insert Dongle, turn on Edge, check LED, re-pair if needed.
Wireless yellow blinking Low battery Stop and charge the Edge.
Wireless switch seems ineffective over USB USB powers the board Disconnect USB before judging switch state.
BLE has no raw EMG BLE mainly provides envelope data Use Dongle serial for raw EMG.

No. Flex is an analog front end. It outputs analog voltage on S. You need an ADC, MCU, or oscilloscope.

electrodes → Flex → ADC / MCU / oscilloscope → serial plotter or browser display
Flex pin Connect to
5V Low-noise 5 V power
G MCU / ADC / oscilloscope ground
S ADC input
VREF / R Optional ADC input for stable analysis

S is about 0–3.3 V, centered near 1.65 V.

Yes. Flex outputs analog voltage. ESP32 is useful for quick validation; external ADC or STM32 ADC + DMA is better for stable sampling.

The left button restarts the Edge. The right button re-pairs it with the Dongle.

The main LED yellow blinking usually means low battery. Stop and charge the Edge.

Does the Type-C blue LED mean fully charged?

Section titled “Does the Type-C blue LED mean fully charged?”

No. It usually means USB / board power is present. Read battery level, charging state, and charging voltage from the console or data interface.

  • The switch must be on for charging.
  • Charging starts about 3 min after USB is connected.
  • A full charge usually lasts about 60–90 min.
  • Avoid wearing the Edge while it is continuously connected to USB.

How many Edge devices can one Dongle connect?

Section titled “How many Edge devices can one Dongle connect?”

Up to 4. For multi-Edge capture, prefer one Dongle for synchronization.

No. It is a USB receiver and must remain plugged into the computer or host.

Both measure the same surface EMG signal. Use dry electrodes for daily exploration, teaching, DIY, and wearable prototypes. Use wet electrodes when you need lower, more stable contact impedance for controlled scientific measurements.

Why is the waveform cleaner away from the laptop?

Section titled “Why is the waveform cleaner away from the laptop?”

Laptops, power adapters, displays, power strips, and USB cables can introduce electromagnetic or ground noise. Use battery power and move away from noise sources for the first test.

  1. Power is present.
  2. Flex S is connected to the ADC input.
  3. G is common with the acquisition device.
  4. Electrodes are attached and IN+ / IN- / REF are not swapped or shorted.
  5. ADC input range covers 0–3.3 V.

See Noise and Flatline for more detail.

Usually Flex. Choose Wireless if you want fewer wires and fast browser-based wireless waveforms. For compact wearable structures, see the Mini product page.

No. AAW Lab products are for teaching, early research, electronics development, and non-diagnostic experiments only.