After Purchase Setup
If you already have the hardware, start here. The first goal is not a perfect waveform; it is a repeatable motion-related EMG response.
Choose your product
Section titled “Choose your product”| Product | Start here |
|---|---|
| AAW sEMG Flex | Flex: wire to ADC / MCU |
| AAW sEMG Wireless | Wireless: power, pair, view waveform |
Success criteria
Section titled “Success criteria”- Relaxed baseline is relatively stable.
- Clenching or wrist motion increases waveform amplitude and energy.
- Releasing returns the waveform to a lower level.
- The waveform is not fixed as a flat line and does not stay saturated.

Flex: wire to ADC / MCU
Section titled “Flex: wire to ADC / MCU”Flex is an analog front end. It must be connected to an ADC, MCU, or oscilloscope.
electrodes → Flex → ADC / MCU / oscilloscope → serial plotter or browser consoleYou need
Section titled “You need”- AAW sEMG Flex.
- Electrode leads with dry electrodes for convenient tests, or wet electrodes for controlled reference measurements.
- ESP32, Arduino, STM32, external ADC, or oscilloscope.
- Battery or low-noise isolated power.
Wiring
Section titled “Wiring”| Flex pin | Connect to | Notes |
|---|---|---|
5V |
Low-noise 5 V |
Battery or isolated power is recommended. |
G |
MCU / ADC / oscilloscope ground | Common ground is required. |
S |
ADC input | EMG analog output, about 0–3.3 V. |
VREF / R |
Optional ADC input | Mid-rail reference, useful for stable analysis. |
First test
Section titled “First test”- Power Flex with low-noise
5 V. - Connect
Sto an ADC input andGto the acquisition ground. - Place
IN+andIN-along the target forearm muscle; placeREFnear the wrist or another quieter location. - Observe raw ADC or oscilloscope data.
- Relax, then clench or bend the wrist and confirm the waveform increases.
Examples
Section titled “Examples”| Goal | Example |
|---|---|
| Arduino analog validation | arduino-analog-teleplot |
| ESP32 analog validation | esp32-analog-teleplot |
| Browser console with Flex | esp32-emg2-serial-monitor |
| Stable STM32 sampling | stm32f103-emg |
Wireless: power, pair, view waveform
Section titled “Wireless: power, pair, view waveform”Wireless uses an Edge device on the skin and a Dongle on the computer.
Buttons and LED
Section titled “Buttons and LED”
| Position | Purpose | Notes |
|---|---|---|
| Switch | Controls Edge battery power | Disconnect USB before judging switch state. |
| Left button | Restart Edge | Usually not needed during normal capture. |
| Right button | Re-pair | Use when reconnecting to the Dongle. |
| Main LED | Status / low battery | Blue means normal operation; yellow blinking usually means low battery. |
First test
Section titled “First test”- Charge the Edge first; a full charge usually lasts about
60–90 min. - Plug the Dongle into the computer.
- Disconnect USB from the Edge and use battery power.
- Turn on the Edge switch.
- Re-pair with the right button if needed.
- Place the Edge onboard electrodes on the target forearm muscle.
- Open console.aawlab.com.
- Relax, then clench or bend the wrist and watch the waveform.
Charging and multi-Edge notes
Section titled “Charging and multi-Edge notes”- The Edge switch must be on for charging.
- Charging starts about
3 minafter USB is connected. - The Type-C blue LED is not a full-charge indicator.
- Up to 4 Edge devices can connect to one Dongle.
- For multiple Edge devices, prefer one Dongle for synchronization.
- BLE mainly provides envelope data; use Dongle serial for raw data.
If there is no waveform
Section titled “If there is no waveform”| Symptom | First checks |
|---|---|
| Flat line | Power, electrodes, ADC input, Flex S wiring. |
| Large noise | Electrode contact, laptop power adapter, cable motion, nearby noise sources. |
| Wireless no data | Dongle inserted, Edge on, pairing, battery level. |
| Flex does not appear over USB | Flex is not a USB acquisition device; use ADC / MCU / oscilloscope. |
See Troubleshooting Table and Noise and Flatline.